BackgroundThis study develops a model-based myocardial T1 mapping technique with sparsity constraints which employs a single-shot inversion-recovery (IR) radial fast low angle shot (FLASH) cardiovascular magnetic resonance (CMR) acquisition. The method should offer high resolution, accuracy, precision and reproducibility.MethodsThe proposed reconstruction estimates myocardial parameter maps directly from undersampled k-space which is continuously measured by IR radial FLASH with a 4 s breathhold and retrospectively sorted based on a cardiac trigger signal. Joint sparsity constraints are imposed on the parameter maps to further improve T1 precision. Validations involved studies of an experimental phantom and 8 healthy adult subjects.ResultsIn comparison to an IR spin-echo reference method, phantom experiments with T1 values ranging from 300 to 1500 ms revealed good accuracy and precision at simulated heart rates between 40 and 100 bpm. In vivo T1 maps achieved better precision and qualitatively better preservation of image features for the proposed method than a real-time CMR approach followed by pixelwise fitting. Apart from good inter-observer reproducibility (0.6% of the mean), in vivo results confirmed good intra-subject reproducibility (1.05% of the mean for intra-scan and 1.17, 1.51% of the means for the two inter-scans, respectively) of the proposed method.ConclusionModel-based reconstructions with sparsity constraints allow for single-shot myocardial T1 maps with high spatial resolution, accuracy, precision and reproducibility within a 4 s breathhold. Clinical trials are warranted.

Highlights

  • This study develops a model-based myocardial T1 mapping technique with sparsity constraints which employs a single-shot inversion-recovery (IR) radial fast low angle shot (FLASH) cardiovascular magnetic resonance (CMR) acquisition

  • We extend a previously developed method [20] for sparsity-constrained model-based T1 estimation to allow for cardiac applications

  • The proposed technique is compared to a reference T1 map obtained by a conventional IR spinecho method

Read more

Summary

Introduction

This study develops a model-based myocardial T1 mapping technique with sparsity constraints which employs a single-shot inversion-recovery (IR) radial fast low angle shot (FLASH) cardiovascular magnetic resonance (CMR) acquisition. Quantitative myocardial T1 mapping finds increasing applications in clinical cardiovascular magnetic resonance (CMR) imaging. Replacing the bSSFP readout by a fast low angle shot (FLASH) acquisition completely avoids banding artifacts [7,8,9,10,11]. Non-Cartesian acquisition schemes (mainly radial) have been employed to enable fast myocardial T1 mapping [9,10,11]. The combination of radial encoding with sliding window image reconstruction [10], compressed sensing [9] and real-time CMR [11] has enabled high-resolution myocardial T1 mapping within a single inversion-recovery (IR) relaxation process

Methods
Results
Discussion
Conclusion