Abstract

A cardiovascular system (CVS) model has previously been validated in simulated cardiac and circulatory disease states. It has also been shown to accurately capture all main hemodynamic trends in a porcine model of pulmonary embolism. In this research, a slightly extended CVS model and parameter identification process are presented and validated in a porcine experiment of positive end-expiratory pressure (PEEP) titrations at different volemic levels. The model is extended to more physiologically represent the separation of venous and arterial circulation. Errors for the identified model are within 5% when re-simulated and compared to clinical data. All identified parameter trends match clinically expected changes. This work represents another clinical validation of the underlying fundamental CVS model, and the methods and approach of using them for cardiovascular diagnosis in critical care.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.