Abstract

Abstract The method of response interpretation for determining dose equivalents of a new multi-element thermoluminescence dosemeter in mixed photon-beta and photon-neutron fields is described. The four TL chips are LiF:Mg,Ti, each covered by its own unique filter. Experiments to determine the response of the dosemeter and to test the dose calibration algorithm were performed according to the standard established by the US Department of Energy Laboratory Accreditation Program (DOELAP). It has been shown that the dosemeter response can be used to identify the mixture type. The accuracy of the system has been proven in blind testing to be well within DOELAP tolerance limits in all test categories.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.