Abstract

A method has been developed to reduce the mass spectrometric ion signal suppression associated with the use of TFA as an additive in LC mobile phases. Through post-column infusion of diluted NH(4)OH solution to LC eluents, the ammonium ion introduced causes the neutral analyte-TFA ion pair to dissociate which consequently releases the protonated analyte as free ions into the gas phase (through regular electrospray ionization mechanisms). An ion signal improvement from 1.2 to 20 times for a variety of compounds had been achieved through the application of this method. The molar ratios of NH(4)OH:TFA which result in a reduction of signal suppression were determined to be between 0.5:1 and 50:1. In addition, it was shown that this NH(4)OH infusion method could reduce the level of doubly-charged species and the product ions formed via in-source collision. The use of diluted NH(4)OH solution is favorable since it is compatible with mass spectrometry analysis, and it is applicable in both positive and negative-ion generation mode.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.