Abstract
We present an 8-year experience with the Wu¨rzburg noncompression titanium miniplate system for the rigid fixation of the mandible during elective head and neck cancer surgery in a consecutive series of 100 patients. One half of the miniplates were used to fix mandibulotomies undertaken for surgical access. The remaining half were in patients undergoing reconstruction of segmental mandibular defects, the vast majority (92%) with vascularised bone grafts.One to four variously shaped miniplates were used per patient (mean = 1.5), plate size ranging from 4 to 40 holes. Fifteen patients (15%) developed complications which included 3 mandibular osteoradionecrosis, 8 broken, 5 infected, and 4 exposed plates. Three of the eight fractured plates were associated with nonunion.In this study, the main advantages of titanium miniplate fixation, namely ease of application, decreased fixation time and malleability, were accompanied by a level of morbidity which, while comparing well with alternatives, may necessitate a reappraisal of this technique of fixation.