Abstract
When tuned for maximum sensitivity, the electron bombardment ion source on the ANU (the Australian National University) Helix MC Plus noble gas mass spectrometer produces an Ar ion beam that is mass discriminated as a function of Ar partial pressure. This effect can be reduced substantially or eliminated by adjusting the repeller and trap voltages, although with a sensitivity loss of up to 50%. A recommended procedure for tuning an ion source to reduce pressure dependent mass discrimination is described. Similar procedures can be applied to other noble gases and other ion sources to ensure that pressure dependent mass discrimination is avoided.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.