Abstract
LAILA is a miniaturized eight-channel electronic readout system for compact γ-ray detectors, combining high-resolution spectroscopy capability with position sensitivity. Compactness is achieved by the combination of a novel CMOS front-end ASIC (Application-Specific Integrated Circuit) for analog processing of a large signal current from Silicon PhotoMultiplier (SiPM) solid-state photodetectors, with a microcontroller-based data acquisition system. The adoption of automatic gain regulation in the gated-integrator stage of the ASIC offers an 84 dB dynamic range, combining single-photon sensitivity with an extended input photon energy range (20 keV-4MeV, using 30 μm-cell SiPMs). Using this module with properly merged 144 SiPM pixels coupled to a 3 in.-thick lanthanum bromide scintillation crystal, a 3% energy resolution at 662 keV and 1cm spatial resolution in the estimation of the interaction coordinates are experimentally demonstrated in this work.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.