A miniature quadrupole mass spectrometer array consisting of 16 rods in a 4×4 array is reported. Each rod is 25 mm in length and 2 mm in diameter. The ionizer is of a miniature Nier-type design, and the detector is a channel-type electron multiplier. Operating frequencies are 5.3, 7.1, and 12.9 MHz. The mass range demonstrated herein is 1–300 u; and the resolution of the system is 0.1–0.5 u (full width at half-maximum), or m/Δm=600. The present sensitivity is calculated and measured to be approximately 1×1012 counts/Torr s.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.