Abstract

High permeability magnetic flux concentrators are used to enhance the sensitivity of an atomic magnetometer operating in the spin-exchange relaxation-free regime. The magnetometer uses a millimeter scale Rb87 vapor cell and either mu-metal or Mn–Zn ferrite flux concentrators. The measured sensitivity gives excellent agreement with calculations of thermal noise from the concentrator material. The mu-metal concentrators allow a sensitivity of 50fTHz−1∕2, limited by thermal current magnetic noise. The ferrite concentrators are limited by thermal magnetization noise at low frequencies, and reach a sensitivity of 10fTHz−1∕2 for frequencies above 125Hz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.