Abstract
A standard of attenuation for microwave measurements has been established using the heterodyne or intermediate-frequency substitution method of calibrating microwave attenuators. A preliminary model of the waveguide-below-cutoff standard attenuator has been built, tested, and improved until its accuracy is within 0.01 decibel per 20 decibels. Accuracy of the entire unit is within 0.02 decibel in the 0–10-decibel range and within 0.2 per cent of the measured attenuation in decibels in the 10–50-decibel range.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.