Abstract

The performance of heart rate (HR) monitoring using wrist-type photoplethysmographic (PPG) signals is strongly influenced by motion artifacts (MAs), since the intensive physical exercises are common in real world. Few works focus on this study so far because of unsatisfying quality of corrupted PPG signals. In this paper, we propose an accurate and efficient strategy, named MICROST, which estimates heart rate based on a mixed approach. The MICROST framework is designed as a MIxed algorithm which consists of acceleration Classification (AC), fiRst-frame prOcessing and heuriStic Tracking. Experimental results using recordings from 12 subjects during fast running and intensive movement showed the average absolute error of heart rate estimation was 2.58 beat per minute (BPM), and the Pearson correlation between the estimates and the ground-truth of heart rate was 0.988. We discuss our approach in real time to face the applications of wearable devices such as smart-watches in reality.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.