Abstract

Reverse-phase supports for the separation of peptides and proteins are compared in two high-performance liquid chromatographic systems. One uses a trifluoroacetic acid-acetonitrile solvent system with a 206-nm detector, and the other uses pyridine-formate or pyridine-acetate and 1-propanol with a postcolumn fluorescence detector. Each system was examined with RP8, RP18, and alkylphenyl supports. In most applications, the trifluoroacetic acid-acetonitrile solvent system used in conjunction with an alkylphenyl column performed best. The use of this system for the preparation of low-microgram amounts of samples for microsequence analysis is illustrated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.