Abstract

A flow-through radioactivity detector was used for the high-performance liquid chromatographic determination of amino acids and other nitrogenous substances labeled with 13N, a short-lived ( t 1 2 9.96 min) positron-emitting radionuclide. 13N-Labeled compounds were analyzed using cation, anion and amino columns, or as the o-phthaldialdehyde derivative on an ODS column. Use of column-switching valves and a high-performance liquid chromatographic system with a quaternary eluting capability permits two to three 20-min analyses of labeled samples from a single 13N experiment to be carried out on different columns using a binary or a single mobile phase. Radioactivity in liver metabolites was quantified using an on-line flow-through monitor with data processing capability for integrating peaks and correcting for radioactivity decay. As an example, 1 min following an l-[ 13N]glutamate injection via the hepatic portal vein, 77% of the label in the liver was in a metabolized form; at least ten labeled products were formed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.