Abstract

This paper describes a selective control of the flow direction in microfluidic chips using a laser induced thermal bubble. The bubble is generated inside the microchannel by focusing a continuous-wave laser onto the tear-drop shaped metal pad underneath the microchannel. Experiments demonstrate that the unidirectionally growing bubble acts as an in-channel directional micro-pump and can drive the flow to a freely chosen direction at a T-junction by pumping at different locations. In addition, the pumping flow rate could be adjusted in the real time and this technology is applicable to many types of common liquids. These advantages make it useful for flexible microfluidic manipulation in lab-on-a-chip systems and microchannel networks.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.