Abstract

Continuous sheathless particle separation with high efficiency is essential for various applications such as biochemical analyses and clinical diagnosis. Here, a novel microfluidic device for highly efficient, sheathless particle separation using an elasticity-dominant non-Newtonian fluid is proposed. Our device consists of two stages: sheathless three-dimensional focusing (1 st stage) and separation (2nd stage). It is designed based on the principle of a viscoelasticity-induced particle lateral migration, which promises precise separation of particles in a microfluidic device. Particles of 5- and 10-μm diameters were all focused at the centerline of a circular channel at the 1st stage and successfully separated at the 2nd stage with an efficiency of ∼99.9% using size-based lateral migration of particles induced by the viscoelasticity of the medium. We also demonstrated the capability of our device for separation of blood cells into multiple fractions. The tunability of separable particle size could be achieved by changing the viscoelastic property of the medium and flow rate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.