Abstract

A microdevice for coulometric detection of organophosphate pesticides (OPs) was fabricated based on the measurement of the inhibition of an enzyme, acetylcholinesterase (AChE), by OPs. Thiocholine (TCh) produced in the enzymatic reaction of AChE with acetylthiocholine (ATCh) as a substrate was oxidized on a microelectrode array formed in a main flow channel. Volumes of plugs of necessary solutions were measured using a structure consisting of a row of rhombuses formed in an auxiliary flow channel. The plugs were merged and solution components were mixed at a T-junction formed with the main and auxiliary flow channels. A linear relationship was confirmed between the generated charge and the logarithm of the OP (malathion) concentration in a concentration range between 10(-6) and 10(-3) M with a correlation coefficient of 0.951. The lower limit of detection was 412 nM.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.