Microfabricated quadrupole mass spectrometers with Brubaker prefilters are demonstrated for the first time. Complete filters are assembled from two dies, each carrying two pairs of rods providing the prefilter and main filter sections. The rods are held in precision silicon mounts that are fabricated using wafer-scale deep reactive-ion etching and anodic bonding to glass substrates. Improvements to ion transmission are obtained by tuning the bias potential applied to the prefilter. The effect is explained in terms of a simple analytic theory for ion motion in the prefilter. Mass filtering with a range of m/z = 0-1200 and a resolution of m/?m ? 150 at 10% of peak height is demonstrated using 2-4-mm-long prefilter electrodes, 30-mm-long main electrodes (both of 650 ?m diameter), and a radio-frequency drive at ?6.5 MHz.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.