Abstract

The simultaneous on-line determination of glucose and lactate using a microdevice that consisted of a dialysis sampling system incorporated to the flow-through cell of a microfabricated biosensor array is presented. The fluidic connections between the different device's components were realized by subsequent processing of stacked dry resist layers on a plastic support that provided also the means for electric connections. The performance of the device was evaluated in vitro. The cross-talk effect on the downstream sensor was investigated and found to be negligible. Recoveries of over 95% for both analytes were achieved when flow rates of the perfusion fluid ≤0.5 μl/min were used. At this flow rate, the response time of the device was 2.4 min, which is acceptable for on-line analysis. The linear response concentration range extended up to 30 mM for glucose and 15 mM for lactate. Interference from electroactive species such as ascorbic acid, 2-acetamidophenol and uric acid, was minimal (less than 5% increase in biosensors signal for all substances tested). In addition, the device presented long-term run stability both in buffer and serum samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.