Abstract

We describe on-line optical measurements of urea concentration during the regular hemodialysis treatment of several patients. The spectral measurements were performed in the effluent dialysate stream after the dialysis membrane using an FTIR spectrometer equipped with a flow-through cell. Spectra were recorded across the 5000-4000 cm<SUP>-1</SUP> (2.0-2.5 micrometers at 1-minute intervals. Optically determined concentrations matched concentrations obtained from standard chemical assays with a root-mean-square error of 0.29 mM for urea (0.8 mg/dl urea nitrogen), 0.03 mM for creatinine, 0.11 mM for lactate, and 0.22 mM for glucose. The observed concentration ranges were 0-11 mM for urea, 0-0.35 mM for creatinine, 0-0.75 mM for lactate, and 9-12.5 mM for glucose.