Abstract

A chemical method for the determination of hyaluronan (hyaluronic acid, HA) has been developed and applied to the human blood plasma. Human blood plasma HA was converted to the ΔDi-HA by digestion with hyaluronidase SD and determined by a sensitive and selective high-performance liquid chromatography (HPLC). The HPLC includes the separation and detection of ΔDi-HA using a graphitized carbon column and fluorometric reaction with 2-cyanoacetamide in an alkaline eluent. The calibration graph for ΔDi-HA was linear over the range 0.2 ng–1 μg. It was revealed that the concentration of HA in normal human blood plasma is very low levels (about 24 ng/ml) in comparison to low-sulfated chondroitin 4-sulfate (about 13 μg/ml).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.