Abstract

A DuPont Moisture Analyzer is used in a microcoulometric method for determining water in minerals. Certain modifications, which include the heating of the sample outside the instrument, protect the system from acid gases and insure the conversion of all hydrogen to water vapor. Moisture analyzer data are compared to concurrent data obtained by a modified Penfield method. In general, there is a positive bias of from 0.1 to 0.2% in the moisture analyzer results and a similarity of bias in minerals of the same kind. Inhomogeneity, sample size, and moisture pick-up are invoked to explain deviations. The method is particularly applicable to small samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.