Abstract
A miniaturized flow cell for an electrical conductivity detector has been made and applied to the ion chromatography of inorganic anions. It consisted of stainless steel tubes (0.13 mm ID×0.31 mm OD) and PTFE tubes (0.25 mm ID×2 mm OD). The detection limit for chloride at S/N=3 was 36 pg or 0.33 ppm for a 0.11 μl injection.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have