Abstract

BackgroundWhen a microcatheter is in the aneurysm, it may move due to low-profile visualized intraluminal support (LVIS) deployment. This study was designed to determine this mechanism. MethodsSix silicon aneurysm models were created by combining the aneurysm location (side wall or bifurcation) and the parent vessel configuration (straight, ipsilateral bending, or contralateral bending). After adjusting the microcatheter tip position in the aneurysm by pushing or pulling, an LVIS stent was deployed to cover the aneurysm neck, and the changes in the microcatheter tip position was measured. Pushing and pulling were performed 15 times each for each model, for a total of 180 experiments. ResultsIn all experiments, the microcatheter tip moved with LVIS deployment. The total movement distance was 3.00±1.59 mm, which was significantly different between the push and pull groups (p = 0.049), between the three side-wall aneurysm models (p<0.0001), and between the three bifurcation aneurysm models (p<0.0001). Backward movement in the aneurysm occurred in 21% (37/180). The frequency of backward movement was significantly different between the side-wall and bifurcation aneurysm models (p = 0.0265) and between the push and pull groups (p<0.0001). The forward movement distance was significantly different between the side-wall (n = 78) and bifurcation (n = 65) aneurysm models (p<0.0001). ConclusionsThe aneurysm location, the parent vessel configuration, and adjustment of the microcatheter tip position by pushing or pulling may affect the total movement distance and forward/backward movement of the microcatheter tip due to LVIS deployment.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call