Abstract

We report measurements of chemical shifts in the Fe-L x-ray lines of different Fe compounds from xray emission spectra acquired using a microcalorimeter energy dispersive spectrometer (EDS). The observed changes in peak position and relative intensity of the Fe-L x-ray lines are in agreement with measurements obtained using a wavelength dispersive spectrometer (WDS), demonstrating the usefulness of microcalorimeter EDS for high-energy-resolution x-ray microanalysis.Chemical shifts result from changes in electron binding energies with the chemical environment of atoms. In x-ray spectra, chemical shifts lead to changes in x-ray peak positions, relative peak intensities, and peak shapes. These chemical bonding effects can be significant (with x-ray peak shifts on the order of 1 eV), particularly for x-ray lines resulting from transitions involving valence electrons of light elements such as B and C.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.