Abstract

A microbiosensor based on an integrated thermopile was designed and fabricated on a quartz chip. The thermopile, which was manufactured by doping boron in polysilicon together with aluminium, provided a potential output of ca. 2 mV K . A silicone rubber membrane was used to form and seal the microchannel. The total column volume was 20 μl. Glucose oxidase and catalase were co-immobilized on spherical CPG beads (controlled pore glass) and in turn charged into the microchannel. Using 1 μl sample volumes, a linear range of 2 to 25 mM glucose was obtained using a flow rate of 105 μl min . The relative standard deviation for 100 glucose samples (10 mM) was 5%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.