Abstract

To compare detection rates of microaneurysms (MAs) on high-speed megahertz optical coherence tomography angiography (MHz-OCTA), fluorescein angiography (FA) and colour fundus photography (CF) in patients with diabetic retinopathy (DR). For this exploratory cross-sectional study, MHz-OCTA data were acquired with a swept-source OCT prototype (A-scan rate: 1.7 MHz), and FA and CF imaging was performed using Optos® California. MA count was manually evaluated on en face MHz-OCTA/FA/CF images within an extended ETDRS grid. Detectability of MAs visible on FA images was evaluated on corresponding MHz-OCTA and CF images. MA distribution and leakage were correlated with detectability on OCTA and CF imaging. 47 eyes with severe DR (n = 12) and proliferative DR (n = 35) were included. MHz-OCTA and CF imaging detected on average 56% and 36% of MAs, respectively. MHz-OCTA detection rate was significantly higher than CF (p < 0.01). The combination of MHz-OCTA and CF leads to an increased detection rate of 70%. There was no statistically significant association between leakage and MA detectability on OCTA (p = 0.13). For CF, the odds of detecting leaking MAs were significantly lower than non-leaking MAs (p = 0.012). Using MHz-OCTA, detection of MAs outside the ETDRS grid was less likely than MAs located within the ETDRS grid (outer ring, p < 0.01; inner ring, p = 0.028). No statistically significant difference between rings was observed for CF measurements. More MAs were detected on MHz-OCTA than on CF imaging. Detection rate was lower for MAs located outside the macular region with MHz-OCTA and for leaking MAs with CF imaging. Combining both non-invasive modalities can improve MA detection.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call