Abstract

For the purpose of performing radial scanning by rotating ultrasound beam in intravascular ultrasound imaging system for diagnosing the state of thrombus and stenosis in a blood vessel such as the coronary arteries of the heart, we are developing micro Coiled Stator UltraSound Motor (CS-USM). Three types of coiled stators with materials of titanium, stainless steel or copper were prepared. These coiled stators were 5 wound acoustic waveguide with 0.5 mm width and 0.1 mm thickness. Size of coiled stator is 0.9 mm diameter and 3 mm length. PZT ceramic vibrator or hydrothermal PZT film vibrator was employed. When two vibrators-type inner rotor CS-USM having stainless steel coiled stator were driven by voltage of 30Vpp at 311 kHz, rotational speed was 3500 rpm, torque was 0.34μNm. Maximum rotational speed of one vibrator-type inner rotor CS-USM having hydrothermal PZT film was 1100 rpm. Maximum rotational speed of one vibrator-type inner rotor CS-USM with titanium or stainless steel coiled stator was 2400 rpm. Rotational speed of the one vibrator-type inner rotor CS-USM with copper coiled stator driven by same conditions was 3200rpm. Maximum rotational speed of two vibrator-type outer rotor CS-USM with stainless steel coiled stator and PZT ceramic vibrators was 1400 rpm.