Abstract

A micro liquid–liquid extraction (MLLE) flask designed for the extraction of bioavailable synthetic pyrethroids from agricultural and ground water samples is used. The method is simple, rapid, and reproducible. Extracts are analyzed off-line by reversed-phase high-performance liquid chromatography (RP-HPLC). A volume of 100 µL n-hexane is sufficient for recovery of the three pyrethroids from 100 mL water. Further clean-up is not required before injection into HPLC. The method is linear in the range of 0.5 to 1000 µg L−1. The method detection limits (MDLs) are in the range of 0.05 to 0.08 µg L−1; recoveries range from 84.7 to 94.5% and enrichment factors from 1058 to 1181. Real surface water samples were analyzed to check the applicability of the proposed method. Cypermethrin (CYP) was detected in agricultural water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.