Abstract

A novel method is proposed to measure NO in breath. Breath NO is a useful diagnostic measure for asthma patients. Due to the low water solubility of NO, existing wet chemical NO measurements are conducted on NO(2) after removal of pre-existing NO(2) and conversion of NO to NO(2). In contrast, this study utilizes direct measurement of NO by wet chemistry. Gaseous NO was collected into an aqueous phase by a honeycomb-patterned microchannel scrubber and reacted with diaminofluorescein-2 (DAF-2). Fluorescence of the product was measured using a miniature detector, comprising a blue light-emitting diode (LED) and a photodiode. The response intensity was found to dramatically increase following addition of NO(2) into the absorbing solution or air sample. By optimizing the conditions, the sensitivity obtained was sufficient to measure parts per billion by volume levels of NO continuously. The system was applied to real analysis of NO in breath, and the effect of coexisting compounds was investigated. The proposed system could successfully measure breath NO.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.