Abstract

<p>Body impedance analysis (BIA) is used to evaluate the human body composition by measuring the resistance and reactance of human tissues with a high-frequency, low-intensity electric current. Nonetheless, the estimation of the body composition is influenced by many factors: body status, environmental conditions, instrumentation, and measurement procedure. This work studies the effect of the connection cables, conductive electrodes, adhesive gel, and BIA device characteristics on the measurement uncertainty. Tests were initially performed on electric circuits with passive elements and on a jelly phantom simulating the body characteristics. Results showed that the cables mainly contribute to increase the error on the resistance measurement, while the electrodes and the adhesive introduce a negligible disturbance on the measurement chain. This paper also proposes a calibration procedure based on a multivariate linear regression to compensate for the systematic error effect of BIA devices.</p>

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.