Abstract

AbstractA gradient high-performance liquid chromatography (HPLC) method has been developed to determine the concentrations of latanoprost (LP) and latanoprost free acid (LPA) in aqueous solutions. It is novel due to a combination of its simplicity, speed, and detection capability in aqueous solutions for both active drug (LPA) and prodrug (LP). This method is applicable for the research and development of novel drug delivery devices and quality control assays for experimental and commercial laboratory settings, as it allows for a high sample throughput. Samples were chromatographed across a C18 + 2.7 µm 4.6 × 7.5 mm reversed-phase column with gradient elution using a mobile phase of aqueous acetic acid (pH 3.1) and acetonitrile with 0.1% acetic acid. UV spectrophotometry was used to monitor the eluents at 210 nm. Drug concentrations from 1.0 to 150 μg mL−1 were tested, with good linearity observed across the range. LPA had a signature peak at approximately 4.82 min (SD < 0.08) and LP at 9.27 min (SD < 0.07). For both drug and pro-drug, LOD and LOQ were 1.0 and 2.5 μg mL−1, respectively. This assay which accurately measures both prodrug and drug in a single injection, has significant applicability in determining the release kinetics of novel LP drug delivery systems.