Abstract
Methodological errors in the precise measurement of γ-ray energies with Ge detectors are examined basing on a model of photopeak shape function. The model revealed that the centroid of the normal distribution embedded in the phenomenal photopeak is to be compared with the energy scale. The peak difference rate is introduced to evaluate the peak difference, the essential quantity of experimentally giving secondary energy standard, relative to the energy difference. Varied value of the gap difference-rate index with several physical parameters has indicated that the secondary γ-ray energies have a definite amount of methodological errors. A typical methodological error in the γ-ray energy at around 300 keV is in the range of −2637±125–-54±123eV (or −877±418–-180±411ppm). Commonly used “conventional” energy-calibration methods by the use of the most probable peak-height position have been reviewed. Origins of methodological errors are discussed. A new energy-calibration method using the centroid is proposed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.