Abstract

The measurement of heart rate during exercise in the horse has been recommended as a means of prescribing work effort, monitoring the changes in aerobic capacity during training and as part of the clinical examination of the performance horse. The accuracy of four heart rate meters (PEH 100, PU 10, HR 14 and HRM-7) was assessed by comparison with heart rate determined by simultaneous telemetry electrocardiography (ECG) using a rapid incremental exercise test on a treadmill. Heart rate displayed by all four meters showed significant correlations with the ECG heart rate, but only two meters (PEH 100 and HRM-7) had correlation coefficients of 0.98 or greater.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.