Abstract

Abstract A method of optical 15N analysis is proposed for application to small amounts of nitrogen gas released from an automatic nitrogen analyzer (model ANA-1300, Carlo Erba, Milano) subjected to certain set modifications. The ANA-1300 was combined with a vacuum line attached by a molecular sieve 13X column. The nitrogen gas released from the ANA-1300 was introduced with a carrier gas of helium into the molecular sieve column which was pre-evacuated at 10-1 Torr and cooled with outer liquid nitrogen. After removal of the helium by evacuation, the nitrogen gas fixed on the molecular sieve was released by warming the column, and then, it was sealed into preevacuated pyrex glass tubes at 4.5–5.0 Torr. In the preparation of discharge tubes, contamination of unlabelled nitrogen occurred from the carrier gas of standard grade helium, and the relative lowering of the 15N value by it was estimated to be less than 1% when over 700 μg nitrogen was charged on the ANA-1300; when 200 μg nitrogen was charged, it was ...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.