Abstract

Krypton-helium mixture was used as an adsorbate for surface area measurement--. The surface area measurements depend on the accuracy with which the krypton concentration is known. Generally gas tanks supplied by Union Carbide provide a nominal value of 0.1% krypton in helium. The surface area measurements require, however, that the krypton concentraion be known to +- 0.001% or better. A standard plot of krypton volume in microliters vs the area under the curve as measured by a planimeter using the helium detector and Molecular Sieve 5A column was obtained. Results with a thermal conductivity detector using Molecular Sieve 5A and Carbon Molecular Sieve are also given. Low levels of krypton in helium can be measured with precision using either a helium or a thermal conductivity detector with Molecular Sieve 5A or Carbon Molecular Sieve columns. 2 figures, 1 table. (DP)