Abstract
A simple, sensitive, economic and specific reverse phase liquid chromatographic method was developed for the simultaneous estimation of Emtricitabine and Ritonavir in bulk drug. Chromatographic conditions consisted of C-18 Column (Shim-pack) 250 x 4.6 mm, particle size 5 µm , mobile phase combination of methanol and water (80:20), flow rate 1ml per minutes, run time 15 minutes and UV detection at 251nm. . The retention time for Emtricitabine and Ritonavir were found to be 3.25 and 7.8 min and average percentage recoveries 99.42% and 99.63% respectively. The validation parameters were found to comply with ICH guidelines. These methods can be further employed in future for the routine determination of Emtricitabine and Ritonavir in bulk drug and formulation.
 Keyword: Emtricitabine, Ritonavir, RP-HPLC, accuracy and linearity.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.