Abstract
A new instrument for the measurement of choroidal blood flow in the fovea is presented. It is based on the laser Doppler method and a confocal optical system with an indirect detection of the Doppler shifted light. The intensity of the laser beam (785 nm) at the cornea is 90 microW. Measurements were obtained from a normal population of 21 subjects under resting conditions without dilating the pupil. The reproducibility of the choroidal blood flow, based on 5 measurements of 10 s each in 5 randomly selected subjects, is 9%. The minimum detectable change for a statistical significance of p < or = 0.05, based on a population of 21 subjects and 10 s measurements, is 9%. This new compact instrument appears to be suitable for the investigation of the physiology and pharmacology of choroidal blood flow and the effect of age-related macular degeneration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.