Abstract

A new liquid membrane, mercury(II) ion-selective electrode (ISE) based on 1,3-diphenylthiourea as a neutral carrier was developed. In solubility tests of 1,3-diphenylthiourea in six different plastizicers, tris(2-ethylhexyl) phosphate (TEHP) showed the best performance and was selected for the membrane. Several ISEs were conditioned and tested for the selection of seven metal ions (Hg2+, Cu2+, Zn2+, Pb2+, Ca2+, Mg2+ and Ag+). The best Nernstian response was obtained for Hg2+ with an experimental slope of 58.6 ± 0.8 mV per decade in the concentration range 6.0 × 10−6–5.0 × 10−4 mol l−1; the presence of the complex Hg(OH)+ ion explains the slope of the response curve. The ISE showed stable detection reproducibility during 6 weeks, with a practical detection limit of 10−6 mol dm−3 and a high selectivity for Cd2+, Pb2+, Co2+, Cu2+, Sr2+, Zn2+, Ca2+, Mg2+, Mn2+, K+, Na+ and NH4+ ions. Only one interference for Ag+ was found. A second linear response curve for Hg2+ was obtained at pH 4 with a slope of 30.8 mV per decade in the concentration range 2.0 × 10−6–2.1 × 10−4 mol l−1.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.