Abstract

An ultrahigh-speed, wide-field OCT system for the imaging of anterior, posterior, and ocular biometers is crucial for obtaining comprehensive ocular parameters and quantifying ocular pathology size. Here, we demonstrate a multi-parametric ophthalmic OCT system with a speed of up to 1 MHz for wide-field imaging of the retina and 50 kHz for anterior chamber and ocular biometric measurement. A spectrum correction algorithm is proposed to ensure the accurate pairing of adjacent A-lines and elevate the A-scan speed from 500 kHz to 1 MHz for retinal imaging. A registration method employing position feedback signals was introduced, reducing pixel offsets between forward and reverse galvanometer scanning by 2.3 times. Experimental validation on glass sheets and the human eye confirms feasibility and efficacy. Meanwhile, we propose a revised formula to determine the “true” fundus size using all-axial length parameters from different fields of view. The efficient algorithms and compact design enhance system compatibility with clinical requirements, showing promise for widespread commercialization.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call