Abstract

The most suitable elevated vacuum (EV) pressure may differ for each individual prosthesis user depending on suspension needs, socket fit, prosthetic components, and health. Mechanical and physiological effects of EV were evaluated in an effort to determine the optimal vacuum pressure for three individuals. Instrumented EV sockets were created based on the participants' regular EV sockets. Inductive distance sensors were embedded into the wall of the socket at select locations to measure limb movement relative to the socket. Each participant conducted an activity protocol while limb movement, limb fluid volume, and user-reported comfort were measured at various socket vacuum pressure settings. Increased socket vacuum pressure resulted in reduced limb-socket displacement for each participant; however, 81-93% of limb movement was eliminated by a vacuum pressure setting of 12 (approximately -9 inHg). Relative limb-socket displacement by sensor location varied for each participant, suggesting distinct differences related to socket fit or residual limb tissue content. The rate of limb fluid volume change and the change in socket comfort did not consistently differ with socket vacuum pressure, suggesting a more complex relationship unique to each individual. Practitioners may use individual responses to optimize socket vacuum pressure settings, balancing mechanical and physiological effects of EV for improved clinical outcomes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.