Abstract

Objective: Prosthetic sockets are custom made for each amputee, yet there are no quantitative tools to determine the appropriateness of socket fit. Ensuring a proper socket fit can have significant effects on the health of residual limb soft tissues and overall function and acceptance of the prosthetic limb. Previous work found that elevated vacuum pressure data can detect movement between the residual limb and the prosthetic socket; however, the correlation between the two was specific to each user. The overall objective of this work is to determine the relationship between elevated vacuum pressure deviations and prosthetic socket fit. Approach: A tension compression machine was used to apply repeated controlled forces onto a residual limb model with sockets of different internal volume. Results: The vacuum pressure-displacement relationship was dependent on socket fit. The vacuum pressure data were sensitive enough to detect differences of 1.5% global volume and can likely detect differences even smaller. Limb motion was reduced as surface area of contact between the limb model and socket was maximized. Innovation: The results suggest that elevated vacuum pressure data provide information to quantify socket fit. Conclusions: This study provides evidence that the use of elevated vacuum pressure data may provide a method for prosthetists to quantify and monitor socket fit. Future studies should investigate the relationship between socket fit, limb motion, and limb health to define optimal socket fit parameters.