Abstract
We used microwave plasma chemical vapor deposition to deposit nanocrystalline and multilayer (nanocrystalline/microcrystalline/nanocrystalline) diamond thin films on Ti-6AI-4V substrates imitating the condyle and fossa components of the temporomandibular joint. We tested the condyle/fossa pairs for wear in a mandibular movement simulator for an equivalent of two years of clinical use. Analysis of the wear surfaces by optical microscopy, scanning electron microscopy (SEM), and Raman spectroscopy showed that damage in both the films was minimal, no loss of film occurred and the wear performance was superior for the multilayer film. Comparisons with an uncoated condyle/fossa pair showed that the coated temporomandibular joint pairs had improved wear performance.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Materials Science: Materials in Medicine
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.