Abstract
The mechanical properties of titanium-alloy aneurysm clips after long-term implantation in the human cranium are unclear. The characteristics of a Yasargil titanium aneurysm clip were evaluated after long-term implantation for 12 years in a patient with a cerebral aneurysm. The closing forces of the retrieved clip before and after implantation were approximately equal. The bending test showed no differences between the retrieved and control clips. Titanium oxide and calcium were identified on the surface of the retrieved clip, which indicated the formation of corrosion-resistant layers. Titanium-alloy clips retain their mechanical properties in the human cranium for a long time.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.