Abstract

Loading from daily activities or from intensive exercise can lead to increased risk of fracture. Implants designed for load bearing purposes, such as repair of articular cartilage and underlying subchondral bone in knees must have the necessary mechanical competence under long term physiological loadings. In this study, the mechanical behaviour of a polymer-based osteochondral scaffold was examined under monotonic and cyclic loading conditions in a phosphate buffered saline solution at 37°C. Monotonic compression tests at selected strain rates were performed in both confined and unconfined conditions to investigate the influence of confinement. The effects of strain rate and sample composition on mechanical properties were also studied. Multi-step cyclic tests were carried out with increasing compressive loads. Changes in secant modulus and residual strain accumulation are monitored. The secant modulus and the number of cycles to failure of the scaffold are obtained and compared with those of human trabecular bone (Topolinski et al., 2011).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.