Abstract
The introduction of new cardiac SPECT cameras has made it practical to do dynamic SPECT imaging and opened the door to performing myocardial blood flow (MBF) imaging with SPECT. In this paper, we describe in detail our approach to dynamic SPECT MBF imaging using a multi-pinhole cardiac SPECT camera and commercially available kinetic analysis software. We use a 1-day rest/stress protocol with 370 MBq injected at rest and 1,000 MBq at stress with a 1- to 2-hour interval between rest and stress imaging. The tracer is injected mechanically over 30 seconds using a syringe pump. Projection data are acquired in listmode for a duration of 11 minutes and then reframed into a dynamic series. Each image is reconstructed independently using vendor-supplied software. The dynamic images are corrected for residual activity and manually corrected for motion using rigid-body translation. The uptake rate constant, K1, is calculated using a 1-tissue-compartment kinetic model and converted to MBF using a previously determined extraction fraction correction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.