Abstract

A technique to distinguish between materials that can and cannot shield microwaves in liquid nitrogen was developed. The samples are placed in a modified X-band waveguide, which operates by the same ‘‘insertion loss’’ principle as a coaxial device used to measure shielding effectiveness from 300 to 1000 MHz. Verification of the waveguide fixture was accomplished by measuring the shielding effectiveness of a good conductor, coppper; a nonconductor, Teflon; and two intermediate conductors, steel and a carbon-filled elastomer. After verification, the waveguide fixture was used at liquid nitrogen temperatures to compare the shielding effectiveness of YBa2Cu3O7−x, a high temperature superconductor, with copper, a known shielding material.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.