Abstract

Background and Objective. Evaluation of agreement, repeatability, and reproducibility of central and minimal corneal thickness (CCT and MCT) measurements obtained by SOCT, the Scheimpflug system, and ultrasound pachymetry. Materials and Methods. 28 eyes of healthy patients were enrolled. Pachymetry measurements were performed with SOCT, the Scheimpflug system, and ultrasound instrument. Each measurement was taken by 3 operators on 3 devices providing a total of 2100 measurements. Results. The mean CCT for SOCT, Scheimpflug system, and ultrasound instrument was 537.92, 545.94, and 555.74 μm, respectively, (P < .001). The respective mean coefficients of repeatability for CCT were 0.61, 0.82 and 0.80, whereas mean coefficients of interoperator reproducibility for CCT were 0.91, 1.11, and 1.25. Conclusions. CCT and MCT measurements show moderate agreement between instruments. The repeatability and interoperator reproducibility of the results obtained by SOCT are somewhat higher. The operator's impact on CCT and MCT measurements is insignificant in all devices.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.