Abstract
A quantitative method that requires only a small volume (50μL) of blood has been developed for the determination of polybrominated diphenyl ethers (PBDEs) and polychlorinated biphenyls (PCBs). Target analytes in both plasma sample (DBSV) and dried blood spot (DBS) were analyzed by a gas chromatography/high resolution mass spectrometer (GC/HRMS). Measurements of standard reference materials by the developed method were in agreement with those certified values. Linear correlation coefficients were found to be 0.9984 and 0.9965 for DBS and DBSV analysis, respectively. Other analytical criteria, such as limits of detection, recoveries, precision, accuracy and linearity of the proposed method are also reported. From recovery studies, the addition of formic acid to the extraction solvent was found to be effective in extracting PBDEs and PCBs from filter paper. The PBDE and PCB levels in spiked DBS were monitored at room temperature for up to 30 days and the variations of target analytes were found to be insignificant. Our results suggest that DBS sampling technique is feasible for PBDE and PCBs biomonitoring in human population.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.