Abstract

The linearity to the number of shower particles observed by a gas proportional chamber was checked by comparison with scintillation counter observation. Linear response up to about 1000 shower particles was obtained. However, the number of shower particles observed by the proportional chamber is approximately a factor 2 higher than that by the scintillation counter, and this difference may be attributed to the different threshold energy between the two counters.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.