Abstract
Flush mounted hot film anemometer probes were used to measure wall shear stress magnitudes on the inside and outside walls of a rigid model of the human aortic arch. The effects of the presence of an Ionescu-Shiley tri-leaflet bioprosthetic heart valve at the entrance of the aortic arch and the side flows through arteries located in the mid-arch region on wall shear stress magnitudes were determined. It was found that the presence of the tri-leaflet valve leads to an elevation of wall shear stress (relative to the same flow without a valve) over the entire aortic arch region by as much as 50 percent. The valve influence extended to about 180 deg from the entrance to the aorta on the inside wall and even further on the outside wall based on extrapolation of available data. Peak wall shear stress magnitudes measured on the outside wall were in the range of 1.5-4.0 N/m2 (15-40 dynes/cm2) over the length of the aortic arch and took on their highest values in the mid-arch region. Inside wall values were of comparable magnitude. It was observed that the presence of the aortic valve and side flow from the top of the aortic arch reduced wall shear stress reversal in the arch region.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.